並前往https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/detail.cfm?mdrfoi__id=13003932&pc=FKX,13003932,12-01-2021,Malfunction,CONCORD%20MANUFACTURING,12-14-2021,FKX,LIBERTY%20SELECT%20CYCLER%20ASSY(NON-VALUATED),Overfill,%22Description%20of%20Event%20or%20Problem:THIS%20IS%20A%20REPORT%20OF%20A%20PATIENT%20WHO%20EXPERIENCED%20AN%20INCREASED%20INTRAPERITONEAL%20VOLUME%20(IIPV)%20EVENT%20COINCIDENT%20WITH%20PERITONEAL%20DIALYSIS%20(PD)%20THERAPY.%20A%20REGISTERED%20NURSE%20(RN)%20REPORTED%20THAT%20DURING%20DWELL%202%20OF%203%20A%20SUPPLY%20BAG%20LINES%20ARE%20BLOCKED%20MESSAGE%20WAS%20NOTED%20ON%20THE%20MACHINE.%20THE%20PATIENT%C2%BFS%20TREATMENT%20WAS%20DISCONTINUED%20AT%20THAT%20TIME.%20THE%20NURSE%20DELETED%20THE%20TREATMENT%20DATA%20TOO%20EARLY%20AND%20PROCEEDED%20TO%20DO%20A%20MANUAL%20DRAIN%20FOR%20THE%20PATIENT.%20THE%20OVERFILL%20EVENT%20WAS%20INDICATED%20DUE%20TO%20THE%20MANUAL%20DRAIN%20BEING%207500ML,%20WHICH%20IS%20313%25%20OF%20THE%20PRESCRIBED%20FILL%20VOLUME%20OF%202400ML.%20AS%20A%20RESULT%20OF%20THE%20IIPV%20EVENT,%20THE%20TECHNICAL%20SUPPORT%20REPRESENTATIVE%20ADVISED%20THE%20RN%20TO%20DISCONTINUE%20USE%20OF%20THE%20CYCLER.%20A%20NEW%20CYCLER%20WAS%20ISSUED.%20THE%20RN%20WILL%20ASSIST%20PATIENT%20WITH%20MANUAL%20PERITONEAL%20DIALYSIS%20THERAPY%20AS%20NEEDED.%20THE%20PATIENT%20RECEIVED%20A%20NEW%20CYCLER%20WHICH%20IS%20WORKING%20WELL.%20THE%20PD%20THERAPY%20IS%20WITHOUT%20ISSUE%20AT%20THIS%20TIME.%20THE%20CYCLER%20WAS%20REPORTED%20TO%20BE%20RETURNED%20TO%20THE%20MANUFACTURER%20FOR%20PHYSICAL%20EVALUATION.%20,%20Additional%20Manufacturer%20Narrative:THE%20PLANT%20INVESTIGATION%20IS%20IN%20PROCESS.%20A%20SUPPLEMENTAL%20MDR%20WILL%20BE%20SUBMITTED%20UPON%20COMPLETION%20OF%20THIS%20ACTIVITY.%20,%20%22